CN | EN
Product Center
product
Glasses ultraviolet transmittance tester

Glasses ultraviolet transmittance tester

 

 

 

 

Glasses Ultraviolet Tester: Real-time Preview Report

Glasses ultraviolet transmittance tester
Real-time preview of test reports and wavelength data
Glasses ultraviolet transmittance tester
Function
Real-time preview of test reports and wavelength data
Archive and call of lens 1nm raw data
Lens transmittance difference △p%, blue light Tsb test
Polarization efficiency P%, photochromic response value
CIE color coordinate positioning (X, Y)
Lab color, △E color difference
Automatic saving and preview of sample test report PDF
Comparison of transmittance values in any band
Custom settings of explicit values of each standard
Selectable 1nm or 5nm test
0.5 second high-speed continuous test, fast viewing of production line
Three-dimensional Lab color icon report output
Standard testing report
Chinese standards
GB 10810.3-2006 Chinese Standard
GB 32166.1-2016 Eye and Face Protection Standard
GB/T 38120-2019 Blue Light Protection Standard
GB 14866-2006 Standard for Protective Glasses
GB/T 11417.5-2012 Contact Lens Standard
Light Industry Standards
QB 2457-99 Sunglasses Standard
QB/T 2506-2017 Optical Lens Standard
QB 2659-2004 Driver's Mirror Standard
International standard
ISO 8980-3 14889(2013)
American Standard
ANSI Z80.3-2018 Sunglasses American Standard
ANSIZ87.1-2020 American standard for protective goggles
ASTM F659-10 Ski goggles American standard
EU standards
BS EN ISO 12312-1:2015 EU standard
EN166~174 standard for protective goggles
Australian Standards
AS/NZS 1067:2016 Sunglasses Standard
AS/NZS 1337.1-2012 Safety Mirror Standard
Canadian Standards
CSA Z94.3-15
*3-in-1 report of European, American, and Australian standards
*Polarized Film Special Report
*Special report for color changing lenses
*Test Report on Step Stained Lens
Instrument technical specifications
Item
Parameter
Wavelength range
280nm~780nm (Raw data saved to 980nm)
Transmittance range
0~100%
Spectral resolution
5nm
Wavelength accuracy
±0.3nm
Integral measurement time
≤1 second
Transmittance indication error
≤1.5%
Transmittance measurement repeatability
≤1%
Q factor indication error
≤0.02
Spot diameter
5mm
Diopter lens
≤±8D (center)
Voltage power
AC210~240V50/60Hz 100VA
High-Speed / Stable
Accurate / Multi-function